The specimen is transferred automatically by basket between the cold chamber and the hot chamber. Small and large capacities.
| Model | TS-162 | TS-340 | TS-500 | TS-1000 |
|---|---|---|---|---|
| Inner dimensions of test room (mm) | 300 × 300 × 250 | 450 × 450 × 360 | 650 × 650 × 500 | 850 × 850 × 700 |
| Volume of test room (mm) | 22 L | 72 L | 211 L | 505 L |
| Load of test room | 20 kg | 30 kg | 50 kg | 60 kg |
| Overall dimensions (mm) | 1560 × 870 × 1545 | 1710 × 1020 × 1845 | 1910 × 1220 × 2265 | 2110 × 1420 × 2665 |
| Pre-heat room: Upper limit preheat temperature | +220 °C | |||
| Pre-heat room: Heating time | Ambient temperature to +200 °C, within 30 minutes | |||
| Pre-cool room: Lower limit pre-cool temperature | -75 °C | |||
| Pre-cool room: Cooling time | Ambient temperature to -70 °C, within 30 minutes | |||
| Test room: High temperature exposure range | Ambient temperature +20 to +200 °C | |||
| Test room: Low temperature exposure range | –65 to -5 °C | |||
| Recovery time | Within 15 minutes | |||
| Temperature fluctuation | ≤±0.5 °C | |||
| Temperature deviation | ≤±3 °C | |||
| Controller | Programmable color LCD touch screen controller | |||
| Cooling system | Mechanical compression refrigeration system | |||
| Safety device | Over-temperature protection; over-current protection; Refrigerant high-pressure protection; Earth leakage protection | |||
| Exterior material | Steel Plate with protective coating | |||
| Interior material | SUS304 stainless steel | |||
| Power Supply | 380 V 50 Hz | |||
| Ambient conditions | 5 °C to +40 °C ≤85% RH | |||